Table 14. Alarm Table
Display
Log
Description
code
code
rE
12
Control probe faulty
or disconnected
SF
27
Configuration not
completed correctly
SrC
35
Maintenance
request
Low and high temperature alarms LO and HI
Table 15. Low and High Temperature Alarm
Par. Description
AHA Absolute high temperature alarm threshold
ALA
Absolute low temperature alarm threshold
30
Delay
Icon
Alarm
(default)
display
relay
-
YES
YES
-
NO
NO
-
YES
YES
Def.
Min
10°C
-100/
-148
2°C
-100/
-148
Buzzer
Reset
YES
Automatic Compressor operation in
NO
Manual
YES
Manual
Max
UOM
User
537/999
°C/ °F
U
537/999
°C/ °F
U
Effects on
control
duty setting mode (par.
c4); dead band OFF
-
-
User terminal
YES
YES