REMOTE ALARM CONNECTIONS
CONNEXIONS DE L' ALARME À DISTANCE
CONTACT RATING 1 AMP MAX CLASS 2 CIRCUIT ONLY
VALEUR NOMINALE DE CONTACT DE 1 AMP MAX. SUR
UN CIRCUIT DE CLASSE 2 UNIQUEMENT
WIRE COLOR
(FIL DE COULEUR)
GREEN (VERT)
BLACK (NOIR)
RED (ROUGE)
To install the remote alarm, make the following
connections:
1. Connect the COMMON terminal on the cabinet switch
to the COMMON wire on the alarm.
2. To get an alarm when the switch contacts open,
connect the OPEN ON FAIL terminal on the cabinet to
the OPEN ON FAIL wire on the alarm.
a. To get an alarm when the switch contacts close,
connect the CLOSE ON FAIL terminal on the
cabinet to the CLOSE ON FAIL wire on the alarm.
The COMMON and CLOSE ON FAIL wires must be
tied together in this application.
3. Plug the alarm system service cord into an electrical
outlet.
The contacts will trip in the event of a power outage, high
temperature alarm or low temperature alarm.
Final Checks
Before start up, be sure to complete the following steps:
1. Make sure that the unit is free of all wood or cardboard
shipping materials, both inside and outside.
2. Check the positions of the shelves, baskets. If you want
to adjust the positions, see instructions in "Shelves",
"Baskets".
3. Verify that the unit is connected to a dedicated circuit.
18
FUNCTION
(FONCTION)
COMMON (COMMUNE)
OPEN ON FAIL
(EN POSITION OUVERTE)
CLOSE ON FAIL
(EN POSITION FERMÉE)