Operating Standards
The units described in this manual are classified for use
as stationary equipment in a Pollution Degree 2 and Over
voltage Category II environment.
These units are designed to operate under the following
environmental conditions:
• Indoor use
• Altitude up to 2000 m (6512 feet).
• Maximum relative humidity 60% for temperatures
from 15°C to 27°C (59°F to 80°F), or maximum relative
humidity 40% for temperatures from 28°C to 32°C
(82°F to 90°F).
• Main supply voltage fluctuations not to drop or exceed by
10% of the nominal voltage.
• Do not connect the unit to a GFCI (Ground Fault Circuit
Interrupter) protected outlet as it may be subject to
nuisance tripping.
• Do not use extension cords with this unit.
Unit Specifications
The data label is located on the left side on top towards
front of the unit.
The specifications of refrigerator like voltage, required wall
breaker amperage and power plug are same for all the
units.
This plug must be plugged into / supplied with its own
individual branch circuit.