Failure
The diaphragm pump is delivering
little or no liquid
Transmitter does not go into
Hold mode
Inductive check-back position indica-
tor not recognized by EasyClean
Screen is blank
Alarm message "Check Levels"
appears on the display
Error message
TEST EEPROM
TEST DISPLAY
TEST CLOCK
ERROR 12C
ERROR: EEPROM
ERROR: CLOCK
ERROR: LCD
ERROR: ACLOW
© 04/07 Mettler-Toledo AG, CH-8606 Greifensee
Printed in Switzerland
Probable cause
Membrane or internal membrane of
check valve is fouled or defective.
Valves of cleaning medium or buffer
are not supplied with power.
Valves of cleaning medium or buffer
are blocked.
Valves of cleaning medium or buffer
are defective.
Medium tubing defective or blocked.
Terminal contact dry.
HOLD or control signal does not trigger.
Inductive check-back position indicator
defective.
Terminal contact dry.
LCD screen defective.
Power supply interrupted.
The flow sensor cannot identify any
media flow during a rinsing, cleaning
and/or calibration cycle.
Terminal contact dry.
Error messages (ERROR) PLC
Probable cause
Self-test interrupted.
CONTROL defective.
CONTROL circuit diagram storage
is defective.
Clock error.
LCD defective.
CONTROL defective.
EasyClean 350 e
Remedy
Remove membrane head. Clean or
replace membranes. See Section 10.2.
Check the wiring as described in
Section 5.3 "Control unit".
Remove valve and clean, possibly
replace.
Replace.
Check hoses for leaks.
Avoid pinching/distortion of tubing.
Check the wiring as described in
Section 5.3 "Control unit".
Check the wiring on the transmitter.
Check whether green LED on auxiliary
print in control unit lights up for 5 sec –
Pause – 2 sec on changeover from "RUN"
to "Service". Possibly replace the print.
Replace inductive check-back position
indicator.
Check the wiring as described in
Section 5.3 "Control unit". Check cable.
Replace the print unit with the PLC.
Check the power supply.
Check flow sensor.
Check water, detergent or buffer valve.
Check pipe.
Check the wiring as described in
Section 5.3 "Control unit". Check cable.
Remedy
Replace circuit board including PLC.
Replace circuit board including PLC.
Replace circuit board including PLC.
Replace circuit board including PLC.
Replace circuit board including PLC.
Replace circuit board including PLC.
EasyClean 350 e
52 500 194