Housing

For the installation of the retractable housing, please refer to the corresponding instruction
manual for the housing.
When using a retractable housing in conjunction with the EasyClean 350 e, it is important to pay
special attention to the following:
– outlet of flushing chamber to be positioned in a upwards direction
This prevents a self drain out of the cleaning agent and buffer and avoids disturbing air bubbles
in the flushing chamber.
NOTICE to the InTrac798 e housing:
Connect the flushing chamber of the InTrac798e
housing as described on the picture. Lead outlet 1
and outlet 2 separately to the waste (don't use
T-pieces).
5.3

Control unit

DANGER
The main connection may only be performed by authorized specialist staff in accordance
with electrical engineering rules.
Please also observe local regulations for the mains connection to the building.
NOTICE
It is strongly recommended to install the "emergency stop switch" (available as an option) or
to provide a mains switch close to the EasyClean unit.
1 Main cable sleeve M16 (main cable by customer)
2 Inductive check-back position indicator housing "Measure", M12
3 Inductive check-back position indicator housing "Service", M12
4 Sleeve control cable transmitter M16
5 Reserve
6 Sleeve alarm light M16 (optional)
7 Sleeve control cable valve unit M16
8 Lead-through for flow sensor M16 (optional)
©
© 04/07 Mettler-Toledo AG, CH-8606 Greifensee
Printed in Switzerland