5.5

Flow sensor

The flow sensor is to be installed at the pump outlet tube.
Procedure:
1. Open valve unit.
2. Lead the cable through the feed-through duct.
3. Attach the sensor to the tube using two cable ties (position side with markings against the
tube).
4. Insert the cable into the spare opening of the control unit.
5. Connect flow sensor according to Section 5.3.
EasyClean 350 e
EasyClean 350 e
52 500 194