Electrical connections of
the automation PCB
Terminal loading
automation PCB
© 04/07 Mettler-Toledo AG, CH-8606 Greifensee
Printed in Switzerland
Connection Function
40
Feed EasyClean 350 e
41
Feed EasyClean 350 e
42
Feed external
43
Feed external
44
Triple alarm light
45
Signal lamp or DCS: "operation"
46
Signal lamp or DCS: "service"
47
Signal lamp or DCS: "alarm"
53
Remote control or DCS*
54
Remote control or DCS*
55
Remote control or DCS*
56
Remote control or DCS*
57
Remote control or DCS*
58
Transmitter wash contact
59
Transmitter wash contact
60
Transmitter control
61
Transmitter control
* Signal duration: 1 second
NOTICE
If an alarm lamp is to be fed by the EasyClean power bloc, the terminals no. 40 and 42
(24 V DC) must be connected to each other as must also no. 41 and 43 (0 V DC).
Designation
Color of
cable
+24 V int.
0 V internal
+24 V
0 V
common
black
L1
brown
L2
yellow
L3
green
B
brown
S
yellow
R
green
K
white
+24 V
black
trigger
brown
trigger
yellow
control
green
control
white
EasyClean 350 e
Connected
to
43 (bridge)
1 (light)
2 (light)
3 (light)
18 (wash)
17 (wash)
7 (control –)
8 (control +)
EasyClean 350 e
52 500 194