en
|
Troubleshooting
4.12 GPS (Global Positioning System,
optional)
NOTICE
Risk of other devices malfunctioning!
If GPS is installed, the device must not be used on aircrafts.
► Do not use the device when on aircraft.
The unit can be equipped at the factory with GPS (Global
Positioning System) for the satellite-based determination of
the device's position.
If the GPS option has been installed, the determined position
data is stored together with the measurement results and
output when the device is read out.
4.13 Printing Results
To print the measurement results, the device can be
connected to the Dräger Mobile Printer BT (see "Order list",
®
page 20) via Bluetooth
.
Observe the Dräger Mobile Printer BT instructions for use.
Alternatively, the measurement results can also be printed
using the Dräger Mobile Printer.
Observe the Dräger Mobile Printer instructions for use.
● In the menu, select Login.
● Enter PIN-1.
● Press
to confirm.
● In the PIN-protected menu, select Printer.
● Under Printing mode, select the Mobile Printer or
Mobile Printer BT.
4.14 Exporting data
The data stored by the device can be exported and then
®
analyzed in Microsoft
Excel, for example. (This function is
available from firmware version 2.0.0).
1. Connect the USB-C cable to the USB-C port of the device
and the USB port of a PC or laptop.
The device turns on.
2. Log in to the PIN-1 menu in the device menu and choose
Transfer mode.
"Transfer mode" is shown on the device display.
3. Open Windows Explorer.
The device is now displayed as a USB drive.
4. Open the path "USB drive\var\scenario_logs\BreathTest".
5. Select the CSV file.
6. Select Copy.
7. Select the desired drive and directory as the storage
location.
8. Select Save.
The CSV file is saved.
9. Turn off the device by pressing the
Transfer mode.
4.15 App connection
The device data can be transferred to the Dräger add app
automatically via Bluetooth.
18
After initial pairing, the Dräger add app connects to the device
automatically and downloads the data. The data are displayed
there as a list, in a ring graphic and also on a map if GPS is
activated. Other data can be added in addition to the
measurement results.
For initial pairing:
1. Install the Dräger add app by following the instructions on
2. Choose the device in the app.
3. Choose Bluetooth in the device menu.
4. Select Bluetooth and choose Pair with app.
5. Confirm the connection in the app.
6. Compare the code on both devices and confirm.
5
An error message is displayed in the event of a device
malfunction. The number below the message is for service
purposes. If the error persists even after turning the device off
and on multiple times, contact authorized personnel, such as
DrägerService.
Error/Message
Device does not
turn on.
Insufficient volume
of breath/
invalid sample.
Device out of tem-
perature range.
↑ ↑ ↑
(Measuring range
exceeded).
Insufficient volume
of breath. [Only for
breath samples
using the mouth-
piece]
Sampling inter-
rupted.
button to exit
Error number.
the screen.
Troubleshooting
Cause
(Rechargeable)
batteries are
empty.
Subject is not
blowing for long
enough or intermit-
tently.
Device is too warm
or too cold.
Measuring range
exceeded. Resid-
ual mouth alcohol
can distort the
measured values.
Volume of breath
sample insufficient.
Irregular exhala-
tion, abrupt cessa-
tion or inhalation.
Device error.
Instructions for use
Remedy
Charge the
rechargeable bat-
tery or replace the
batteries.
Blow more strongly
and without inter-
ruption into the
mouthpiece/funnel.
If necessary, take a
manual or passive
breath sample.
Allow the device to
cool down or warm
up.
A wait time of at
least 15 minutes
after alcohol was
last consumed
must have
elapsed. To repeat,
press the
but-
ton.
Repeat the breath
sample after press-
ing the
button.
If necessary, take a
manual or passive
breath sample.
Repeat the breath
sample after press-
ing the
button.
If necessary, take a
manual or passive
breath sample.
Turn the device off
and then back on.
If the error persists,
contact DrägerSer-
vice or your dealer.
®
|
Dräger Alcotest