PROGRAMMING
Push and quickly release CAL button and scroll the menu using + and – buttons
till the display shows "Function Type". Push CAL to enter the submenu and
then by pressing + and – button choice " Function mA Input" and confirm
pressing CAL.
Push + and – button to choose the minimum current (mA Min) and confirm by
CAL .
Push + and – button to choose the percentuage of dosing frequency (Imp Min)
of the pump when the input current is equal to mA Min. Confirm by CAL .
Pushing + or – button choice if, when the current is lower than mA Min, the
pump has to dose (Yes) or not (No). Confirm by CAL .
Push + and – button to choose the maximum current (mA Max) and confirm by
CAL .
Push + and – button to choose the percentuage of dosing frequency (Imp Max)
of the pump when the input current is equal to mA Max. Confirm by CAL.
Pushing + or – button choice if, when the current is higher than mA Max, the
pump has to dose (Yes) or not (No). Confirm by CAL .
Push ESC button to esc the menu and go back in measure with the Pulse
functioning just programmed.
GRAPH OF THE mA OPERATION