ITC DOSITEC MF Manuel D'installation Et D'entretien page 43

Publicité

Les langues disponibles

Les langues disponibles

STAND BY (STOP)
MENU TREE
LANGUAGE SELECTION
DELAY START
Rev 3.0
Push and hold the ESC/SBY button for 2 seconds during the functioning to
put the pump in stand by mode:
1) It stops dosing
2) Green on led blinks.
3a) If the functioning is Manual or mA Input the display shows the measure
of mA input on the second row and on the first row it shows alternativenly
the writing "Stop " and the value of flow rate % : "Man.xx%" (for
Manual functioning) and "Off xxx%" (for mA functioning) ; see figure on
the left side.
3b) On the second row the display shows the value of the programmed n if
the functioning is Pulse 1xn or 1xn(M) and the number of pulses to be
received from the water-meter if the functioning is 1/n. On the first row
it shows alternativenly the writing "Stop " and the programmed mode
of functioning see figure on the left side.
Push again the ESC/SBY button for 2 seconds to return the pump to the
programmed functioning.
Push and quickly release CAL button and then pressing + and – buttons select the language
choice. Push CAL to enter the language submenu and select "Language English" by pressing
+ or - .
Push CAL to confirm the choice and ESC to go back in measuring.
Push and quickly release CAL button during the
functioning to enter into the programming menu.
Push + or - button to navigate through the menu.
Push the CAL button to enter the submenus.
WHAT IS IT? The delay start is the time in seconds (0-
999sec) after the power on that the pump waits before
dosing the chemical product. During this period of time the
display shows the message "Delay" on the first row. The
second row shows the value of mA input if the pump is
functioning in manual mode or mA and the programmed
value of n if the pump is functioning in Pulse mode (see
figure on the left side). During the delay start the pump can't
dose but it is possible to enter the menu and modify the
parameters and calibrations.
Cod. 00.009.307
43