Speed control
VACSTAR vacuum pump with VC 10 lite/pro vacuum controller
Automatic adjustment of the "analog speed control" mode if the
analogue connection cable (Pos. 8) is connected to the pump with
the vacuum controller.
Once the target value is reached the pump speed is automatically
reduced to a minimum "0 rpm". Once the target value is reached,
Illustration speed control - Vacuum pump VACSTAR with vacuum Controller VC 10 lite/pro
Check whether the voltage given on the type plate corresponds to
the available mains voltage.
If these conditions have been met, the machine is ready for opera-
tion when the mains plug is plugged in.
Start screen layout at the time of delivery
36
Commissioning
The socket used must be earthed
(fitted with earth contact).
the pump operates according to the leakage rate of the vacuum
system.
Automatic boiling point recognition is only possible with the
VC 10 pro .
If these conditions are not met, safe operation is not guaranteed
and the machine could be damaged.
Also pay attention to the ambient conditions listed in the "Tech-
nical data".
The start screen appears for a few
NOTICE
seconds when the device is swit-
ched using the "Standby ON/OFF"
switch (F) The device name and the
software/firmware
also displayed.
An information prompt then appears to
download the firmware update tool.
versions
are