Periodic Inspection and Maintenance

This unit requires regular inspection. The frequency will be
determined by the operating conditions, which includes extreme
temperatures, exposure to contaminants, or gas. Inspect the unit at
least once every 6 months.
Contact a Honeywell Sales or Service representative for information
about maintenance. Calibration kits with instructions are available
from Honeywell.

Maintenance

The unit will provide years of service with minimal care.
Visually inspect at regular intervals to ensure optimum operating
condition (no breakage, sensor filter not blocked or clogged,
etc.).
An accurate maintenance log of all maintenance, calibration,
and occurrences must be kept for the proper service of this
product.
Do not expose the sensor to high pressure water spray. Sensors
should not be exposed to solvents.
Clean the exterior with a soft, damp cloth. Do not use solvents,
soaps, or polishes.
Do not immerse the unit in liquids.

Sensor Life Span

Sensor life span may be affected by certain operating conditions or by
exposure to concentrations higher than the detection range.
Sensor Type
CO
H
S
2
NO
2
O
2
COMB
3
E
Point Remote Gas Sensor User Manual
Life Span Specifications (typical)
6 years, normal use at temperatures > -10°C
2 years, in air
2 years, in air
2 years until readings are at 85% of original 20.9% input
2 years, in air
Periodic Inspection and Maintenance