gefran 1000 Manuel D'utilisation page 14

Masquer les pouces Voir aussi pour 1000:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 26
Configuration Phase 2 (CFG2)
Access to phase CFG2 is conditioned by the closing of jumper 3 / J13 on the CPU
board on components side (see Hardware Configuration).
Pro/: Software protection level in range 0-31 as per the table.
Pro Setpoint
Allarms
0
DISPLAY <>
DISPLAY <> DISPLAY <> DISPLAY <> DISPLAY <>
1
DISPLAY <>
DISPLAY <> DISPLAY <> DISPLAY <>
2
DISPLAY <>
DISPLAY <> DISPLAY <> DISPLAY <>
3
DISPLAY <>
DISPLAY <> DISPLAY <>
4
DISPLAY <>
DISPLAY <> DISPLAY <>
5
DISPLAY <>
DISPLAY <>
6
DISPLAY <>
DISPLAY
7
DISPLAY
DISPLAY
DISPLAY = DISPLAY
+8 to disable the ON-OFF function from the panel keys
+16 to inhibit the AUTO/MAN function (only for instruments with AUTO/MAN function):
only power display in Automatic remains displayed.
Standard protection level 19. Phase CFG2 parameters are not subject to software
protection (access to phase CFG2 depends only on state of jumper 3/J13(HW
config.).
_AI/: Alarm output function.
Configuration of symmetrical deviation alarms requires setting of positive setpoints
only. For correct function, negative values are not allowed (even if settable).
0 = Alarms 1 and 2 direct absolute (relay energized when alarm limit is exceeded).
1 = Alarm 1 deviation, Alarm 2 absolute, both direct.
2 = Alarm 1 absolute, Alarm 2 deviation, both direct.
3 = Alarm 1, Alarm 2 direct deviation.
4 = Alarm 1 reverse absolute (relay energized below alarm limit),
Alarm 2 direct absolute.
5 = Alarm 1 reverse deviation, Alarm 2 direct absolute.
6 = Alarm 1 reverse absolute, Alarm 2 direct deviation.
7 = Alarm 1 reverse deviation, Alarm 2 direct deviation.
8 = Alarm 1 direct absolute, Alarm 2 reverse absolute.
9 = Alarm 1 direct deviation, Alarm 2 reverse absolute
10 = Alarm 1 direct absolute, Alarm 2 reverse deviation.
11 = Alarm 1 direct deviation, Alarm 2 reverse deviation.
12 = Alarm 1, Alarm 2 reverse absolute.
13 = Alarm 1 reverse deviation, Alarm 2 reverse absolute.
14 = Alarm 1 reverse absolute, Alarm 2 reverse deviation.
15 = Alarm 1, Alarm 2 reverse deviation.
By adding 16 to the selected function code (ex. 9+16 setting 25), AL1 becomes sym-
metrical deviation: in this case, a direct alarm corresponds to an alarm with relay ener-
gized outside the "window", while a reverse alarm corresponds to an alarm with relay
energized inside.
By adding 32, AL2 becomes symmetrical deviation.
By adding 48, both alarms become symmetrical deviation.
Out/: Main output function (Heat/Cool) and AL1 (PD) enable HB alarm and selection
of temperature scale (°C or °F)
Set code for selected combination of functions as per the table..
Nr.
Al.HB
AL1
OUT
0
Disabled
NO PD
HEAT
1
Enabled
NO PD
HEAT
2
Disabled
PD
HEAT
3
Enabled
PD
HEAT
4
Disabled
NO PD
COOL
5
Enabled
NO PD
COOL
6
Disabled
PD
COOL
7
Enabled
PD
COOL
By adding the following 6 STEPS to the 16 combinations, you can get various combi-
nations for AL3.
+0
AL3 NORMAL DIRECT ABSOLUTE
+16
AL3 NORMAL DIRECT DEVIATION
+32
AL3 NORMAL REVERSE ABSOLUTE
+48
AL3 NORMAL REVERSE DEVIATION
+80
AL3 DIRECT DEVIATION WINDOW*
+112 AL3 REVERSE DEVIATION WINDOW*
When limit AL3 = 0 or is negative, AL3 remains energized or de-energized at all times.
Note: The PD alarm cannot be symmetrical.
7 • PROGRAMMING
Phase 0
It, dt
CFG1
<> = SETTING
Scale
Nr.
Al.HB
AL1
OUT
°C
8
Disabled
NO PD HEAT
°C
9
Enabled
NO PD HEAT
°C
10
Disabled
PD
HEAT
°C
11
Enabled
PD
HEAT
°C
12
Disabled
NO PD HEAT
°C
13
Enabled
NO PD HEAT
°C
14
Disabled
PD
HEAT
°C
15
Enabled
PD
HEAT
TyP/ Type of input probe or linear scale.
Cod.
1000
0
J 0...800°C/32...999°F
1
K 0...999°C/32...999°F
2
N 0...999°C/32...999°F
3
S 0...999°C/32...999°F
4
R 0...999°C/32...999°F
5
T -100...400°C/-148...752°F
6
PT100 -199...400°C/-199...752°F
7
PT100 -19,9...99,9°C/-19,9...99,9°F
8
0-50mV no decimals (xxx)
9
0-50mV 1 decimals (xx.x)
10
0-50mV 2 decimals (x.xx)
11*
PT100 -19,9...99,9°C/-3,8...99,9°F
12
10-50mV no decimals (xxx)
13
10-50mV 1 decimals (xx.x)
14
10-50mV 2 decimals (x.xx)
* Hardware version for RTD input with double resolution for greater stability of tenth of
degree. This version does not allow reconfiguration of the input type. For TC probes type S
and R, precision is in the instrument class (0.5%) for temperatures >200°C.
Ct.A / AL1 Cycle time in range 0...200 sec.
dt.A / AL1 Derivative time in range 0.00...9.99 min. (1000)
0.00...19.99 min (1001-1101).
oFt / Input offset adjustment
for 1001/1101:
-300/300
for type 0,1,2,3,4,5,6,8
-30.0/30.0 for type 7,9,11
-3.00/3.00 for type 10
for 1000:
-199/300
for type 0,1,2,3,4,5,6,8
-19.9/30.0 for type 7,9,11
-1.99/3.00 for type 10
LO.S / Minimum value of setpoint and absolute alarms in scale range of selected
probe. Start of linear scale for probe type 8, 9, 10, 12, 13, 14.
Hl.S / Maximum value of setpoint and absolute alarms in scale range of selected
probe. End of linear scale for probe type 8,9,10, 12, 13, 14.
rEL/ Setting of alarm output state in case of broken probe (message Sbr/Err on dis-
play), according to table:
rEL
Output AL1
0
ON
1
ON
2
OFF
3
OFF
4
ON
5
ON
Scale
6
OFF
°F
7
OFF
°F
Note: In case of broken probe, each relay assumes the set state (ON = energized,
°F
OFF = de-energized), which does not depend on the type of alarm (direct or reverse).
°F
A.r.F/ Selection of alarm output function.
°F
Let you assign one of the following functions to each alarm output: Normal Alarm, HB
Alarm, LBA Alarm, Alarm disabled (OFF logic state).
°F
Choose one of the 63 combinations according to the table:
°F
Ar.F
Output AL1
Output AL2
°F
0
AL1
1
HB
2
LBA
3
OFF
4
AL1
5
HB
6
LBA
7
OFF
By setting +0 output AL3 has function AL3
By setting +16 output AL3 has function HB
By setting +32 output AL3 has function LBA
By setting +48 output AL3 is always OFF
14
1001/1101
J 0...800°C/32...1472°F
K 0...1300°C/32...1999°F
N 0...1300°C/32...1999°F
S 0...1600°C/32...1999°F
R 0...1600°C/32...1999°F
T -100...400°C/-148...752°F
PT100 -200...400°C/-328...752°F
PT100 -199,9..199,9°C/ -199,9..199,9°F
0-50mV no decimals (xxxx)
0-50mV 1 decimals (xxx.x)
0-50mV 2 decimals (xx.xx)
PT100 -19,9...199,9°C/-3,8...199,9°F
10-50mV no decimals (xxxx)
10-50mV 1 decimals (xxx.x)
10-50mV 2 decimals (xx.xx)
Output AL2
OUTPUT AL3
ON
ON
ON
OFF
ON
ON
ON
OFF
OFF
ON
OFF
OFF
OFF
ON
OFF
OFF
Ar.F
Output AL1
Output AL2
AL2
8
AL1
AL2
9
HB
AL2
10
LBA
AL2
11
OFF
HB
12
AL1
HB
13
HB
HB
14
LBA
HB
15
OFF
LBA
LBA
LBA
LBA
OFF
OFF
OFF
OFF

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

10011101

Table des Matières