Checking Functionality Cmm Easy-Charge System With Remote Setpoint 0-24 V; Checking Functionality Of The Cmm Easy With External Control Kit - ITW Simco CMM Easy Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Chart 2, settings of the CMM Easy-charge system with External Setpoint 0-10 V
Remote setpoint
0-10 V DC
0 V
2 V
4 V
6 V
8 V
10 V

6.3 Checking Functionality CMM Easy-charge system with Remote Setpoint 0-24 V

- Supply 0 V DC at the "Remote"-input of the Control Kit (switch S1 open).
- Switch on the mains voltage (24 V DC).The LED indicators on the generator will blink quickly
with a green light during start-up. They will then blink slowly as long as no high voltage
setpoint has been selected. "Cycle OK"-output is not active.
- Supply the desired setpoint on the 0-24 V-input of the setpoint connector on the Control Kit.
The setpoint chosen can be checked by measuring on the "Setpoint Feedback"-connector
(see chart 3).
- The high voltage output of the generator will be activated if 24 V DC is supplied at the
"Remote"-input (switch S1 closed).
- Check the manual of the CMM Easy for additional details about the calibration and operation
of the CMM Easy-charge system.
Chart 3, settings CMM Easy-charge system with External Setpoint 0-24 V
Remote setpoint
0-24 V DC
0 V
4,8 V
9,6 V
14,4 V
19,2 V
24 V

7 Checking functionality of the CMM Easy with External Control Kit

The system is working properly when the LED indicator on the External Control Kit lights up at
every charging pulse. If necessary, use the "Setpoint feedback"-signal to check the setpoint
chosen. Check the CMM Easy manual for more details on the operation of the CMM Easy-
charge system.
See chapter 9 (Faults) to solve any problems.
ExternalControlKit_CMMEasy_UM_9752098550_NL_D_GB_F_V2_0
Setpoint
Feedback-output
Control Kit
0 V
2 V
4 V
6 V
8 V
10 V
Setpoint
Feedback-output
Control Kit
0 V
2 V
4 V
6 V
8 V
10 V
Setpoint output
for CMM Easy
0 mA
4 mA
8 mA
12 mA
16 mA
20 mA
Setpoint output
for CMM Easy
0 mA
4 mA
8 mA
12 mA
16 mA
20 mA
High voltage
output
CMM Easy
0 kV
0 kV
-4,5 kV
-9 kV
-13,5 kV
-18 kV
High voltage
output
CMM Easy
0 kV
0 kV
-4,5 kV
-9 kV
-13,5 kV
-18 kV