where:
<STX>=
<ADDR> =
<ADDR> =
<WINDOWS>= ' 000 '...' 999' window number the meaning of the
<COMMAND>= 0x30: window value reading
<DATA> =
<ETX>=
<CRC>=
When a slave device is addressed by the master:
1
Turbo-V 700 ICE Controller User Manual / 87-900-904-01 (L)
0x02
0x80 (for RS 232 and RS 422 only)
0x80 + device number (0...31)
0xFF: brodcasting command (recognized by all the
devices, it doesn't implicate any answer)
(for RS 485 only)
window depends to the device type
0x31: window writing
alphanumeric ASCII string containing, in the case of
writing operation,the parameter to input into the
window addressed by the field <WINDOW>This field
may have variable length according to the data type
contained in the window where you are working in. In
the case of reading request of a window, the data field
doesn't exist.
0x03
XOR among all the characters following
<STX>=(withexception of <STX>), including the end
character <ETX> hexadecimally encoded by twoASCII
characters.
In case of reading request of the value contained in a window, the
slave answers a string equal to the one sent by the master but in
addition there is the field <DATA> containing the value of the
window. The format of the field <DATA> depends to the window
type.
Technical Information
Optional Serial Port
225/262
17