All the logic output from the controller must be connected at P2
mating connector. With the optional P2 mating connector make the
connection with AWG 24 (0.25 mm
indicated in the figure to obtain the desired capability. The following
table describes the signals available on the connector.
The following figure shows a typical logic output connection (relay
coil) but any other device may be connected e.g. a LED, a computer,
etc., and the related simplified circuit of the controller.
Figure 12
Tab. 4
PIN
1-10 *
2-10∗
Turbo-V 700 ICE Controller User Manual / 87-900-904-01 (L)
15
9
Typical output connection open collector
DESCRIPTION
Normal S1 signal 24 V 60 mA, optically isolated output (pin 10 positive, pin 1
negative). The output Voltage will be present when the rotational speed of the
pump is higher than the selected high speed threshold.
Normal S2 signal 24 V 60 mA, optically isolated output (pin 10 positive, pin 2
negative). The output Voltage will be present when the rotational speed of the
pump is higher than the selected low speed threshold.
Technical Information
2
) or smaller wire to the pins
8
8
15
9
1
1
17
Interconnections
+ 24V
215/262