Instructions for Use
Error Messages
MESSAGE
FAULT: SHORT
CIRCUIT
OVERVOLTAGE
FLOW METER
ALARM
204/262
DESCRIPTION
In normal operation (after the
starting phase) the output
connection is shorted (output
current higher than
16 A).
A failure has occurred in the
controller power supply section,
or the controller has received a
spurious signal.
Alarm relating to the purge flow.
The flow value remained at less
than the set threshold for 10
seconds or more.
Turbo-V 700 ICE Controller User Manual / 87-900-904-01 (L)
REPAIR ACTION
Check connections and
shortages between pump
and controller. Press the
START push-button twice
to start the pump.
Press the START button
twice to start the pump
again. If the message is
redisplayed, contact
Agilent for maintenance.
Check the correct
operation of the gas supply
circuit.