Calibration
To enter the "calibration" mode, press any key at the same
time as switching on the instrument.
Press the
A number between 5,000 and 50,000 is displayed.
If this is not the case, the scale must be re-calibrated.
If the display is stable, press the
zero.
Then place the calibration weight (3/6 kg = 500 g,
30 kg = 1000 g) on the platform.
(display 200000 ... 800000 at max. weight).
Enter the value of the calibration weight (3/6 kg = 500 ,
30 kg = 1) and press the
Switch off the scale, remove the load and switch the scale
on again.
The scale is ready for use.

Error messages

Error
Max. weighing range
exceeded.
Weighing platform not cor-
rectly mounted. Scale outsi-
de set-to-zero range.
Set to zero not possible.
Battery almost empty.
key for scale A or the
key to set the scale to
key to calibrate the scale.
Remedy
Remove weighing product
from scale!
Mount platform correctly!
Make sure that the scale is
empty!
Connect the power plug!