Télécharger Imprimer la page

Hach 9610sc SiO2 Fonctionnement page 16

Masquer les pouces Voir aussi pour 9610sc SiO2:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 64
Option
RUN DAYS
START TIME
INTERVAL
DURATION
OFF DELAY
7. Select WARNING in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
WARNING LEVEL
8. Select PROCESS EVENT in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Note: More than one option can be selected.
Option
MEASURING 1
MEASURING 2
MEASURING 3
MEASURING 4
MEASURING 5
MEASURING 6
ZERO CAL
SLOPE CAL
SHUTDOWN
STARTUP
GRAB SAMPLE
MARK END OF MEASUR
Set the error hold mode
1. Push menu and select SETUP SYSTEM>SETUP OUTPUTS>ERROR HOLD MODE.
2. Select an option.
Option
HOLD OUTPUTS
TRANSFER OUTPUTS Switches to the transfer mode when communications are lost. The outputs transfer
16 English
Description
Selects the days the relay stays on. Options: Sunday, Monday, Tuesday, Wednesday,
Thursday, Friday, Saturday
Sets the start time.
Sets the time between activation cycles (default = 5 min).
Sets the period of time the relay is set to on (default = 30 sec).
Sets the time for additional hold/output time after the relay is set to off.
Description
Sets the level for the warning activation and starts applicable individual warnings.
Description
Closes relay during measurement cycle.
Closes relay during measurement cycle.
Closes relay during measurement cycle.
Closes relay during measurement cycle.
Closes relay during measurement cycle.
Closes relay during measurement cycle.
Closes relay during ZERO CAL cycle.
Closes relay during SLOPE CAL cycle.
Closes relay when in SHUTDOWN mode.
Closes relay during STARTUP cycle.
Closes relay during GRAB SAMPLE measurement.
Closes relay for 1 second at the end of each measurement cycle.
Description
Holds the outputs at the last known value when communications are lost.
to a pre-defined value.

Publicité

loading