Télécharger Imprimer la page

Hach 9610sc SiO2 Fonctionnement page 13

Masquer les pouces Voir aussi pour 9610sc SiO2:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 64
Configure the outputs
4–20 mA module setup
1. Push menu and select SETUP SYSTEM>SETUP OUTPUTS>4–20mA SETUP.
2. Select the OUTPUT.
3. Select an option.
Option
ACTIVATION
SELECT SOURCE Select the output. Options: None if the output is not configured, the analyzer name or
SET PARAMETER Select the measurement channel from the list.
SET FUNCTION
SET TRANSFER
SET FILTER
SCALE 0mA/4mA
4–20 mA activation options
1. Push menu and select SETUP SYSTEM>SETUP OUTPUTS>4–20mA SETUP.
2. Select the applicable OUTPUT.
3. Select SET FUNCTION>LINEAR CONTROL and then select the applicable options in the
ACTIVATION menu.
Option
SET LOW VALUE
SET HIGH VALUE
4. Select SET FUNCTION>PID CONTROL and then select the applicable options in the
ACTIVATION menu.
Option
Description
SET MODE
AUTO—the signal is automatically controlled by the algorithm when the analyzer uses
proportional, integral and derivative inputs.
MANUAL—the signal is controlled by the user. To change the signal manually, change the
% value in MANUAL OUTPUT.
PHASE
Selects the signal result when process changes occur.
DIRECT—signal increases as the process increases.
REVERSE—signal increases as the process decreases.
SET SETPOINT Sets the value for a control point in the process.
PROP BAND
Sets the value for the difference between the measured signal and the necessary setpoint.
INTEGRAL
Sets the period of time from the reagent injection point to the contact with the measuring
device.
Description
The menu list items change with the selected function. Refer to
options
on page 13 for more information.
calculation if a calculation formula has been configured. Refer to
on page 12.
Select a function. Further options will vary depending on which function is chosen.
LINEAR CONTROL—Signal is linearly dependent on the process value. PID CONTROL
—Signal works as a PID (Proportional, Integral, Derivative) controller. LOGARITHMIC—
Signal is represented logarithmically within the process variable range. BILINEAR—
Signal is represented as two linear segments within the process variable range.
If TRANSFER is or will be selected as the ERROR HOLD MODE, select SET
TRANSFER and enter the transfer value. Range: 3.0 to 23.0 mA (default = 4.000). Refer
to
Set the error hold mode
Enter the filter value. This is a time-average filter value of 0 to 120 seconds (default = 0).
Select the scale (0–20 mA or 4–20 mA).
Description
Sets the low endpoint of the process variable range.
Sets the high endpoint (upper value) of the process variable range.
on page 16.
4–20 mA activation
Set up the calculation
English 13

Publicité

loading