• Once the desired axes have been selected, the UP arrow
X-Axis Calibration
1. Place the unit into calibration mode as discussed above.
2. Position the unit so that the X axis is directed to the two walls.
3. Press the X/Y
calibration control to the X axis.
4. Press UP arrow and DOWN arrow buttons, conversely to adjust
the laser height to coincide with the zero position of the target.
The zero position is half way between X= and X- at the long shot.
5. Press the ENTER button on the remote control to accept the
calibration value, noting that the X LED goes "OFF".
6. After calibration is completed, make sure to power off the unit
and then power on again to activate the calibration.
Y-Axis Calibration
1. Place the unit into calibration mode as discussed above.
2. Position the unit so that the X axis is directed to the two walls.
3. Press the X/Y
calibration control to the Y axis.
4. Press UP arrow and DOWN arrow buttons, conversely to adjust
the laser height to coincide with the zero position of the target.
The zero position is half way between Y= and Y- at the long shot.