Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS 3
2
Technical Description
2.1
Technical Data
Nominal line voltage:
Permissible line voltage fluctuation:
Nominal current:
Nominal frequency:
Power line resistance:
Fuse at the distribution panel:
Rating:
Tube voltage:
Tube current:
Curve form of high voltage:
Rotation time:
Exposure time:
Reproduction scale:
Focus size, according to IEC 336,
measured in central ray:
Focus marking:
Automatic exposure blockage (see page 27):
Equipment of protective class I
Protection against electric shock:
Protection against penetration of water:
Year of manufacture
Mode of operation:
Long time power rating:
60 82 262 D 3285
D 3285.201.04.06.02
All manuals and user guides at all-guides.com
208V / 230V - 240V
±10%
max 9.7A
50/60Hz
max. 0.8Ohm
20A slow blow
2.1kW
60 – 80kV
6 –11mA
high frequency
Multipuls Residual ripple ≤ 4kV
see page 30
see page 30
With P1 program, medium mandibular arch (plane center)
ca. 1:1.19. The image at the image receiver is approxi-
mately 19% larger than the real proportions.
0.5mm
The duration of the exposure blockage (cool-off period)
depends on the kV/mA step set and the actually triggered
radiation time. Depending on the tube load, pause times up
to 200s are set automatically.
Example: For P1 program with exposure data 74kV/10mA
and a radiation time of 11.3s a pause of 200s results.
Type B equipment
Ordinary equipment (without protection)
Continuous operation
64W
2 Technical Description
(on the name plate)
9