Errors
Error Type
HTM10 error
GM10 error
Internal error
16
All manuals and user guides at all-guides.com
Cause and Possible Solution
Problem with HTM10 module.
- Check that the module sits firmly
in place. Remove and reconnect.
- Check for missing or damaged
HUMICAP® sensor.
- Check for condensation on the
HUMICAP® sensor. Wait for the
sensor to dry out.
- Replace the module if unable to
remove the problem.
Problem with GM10 module.
- Check that the module sits firmly
in place. Remove and reconnect.
- Check that supply voltage is in
range.
- Replace the module if unable to
remove the problem.
Internal problem with the transmitter.
- Restart the transmitter.
- Restore the factory settings using
service port if reset does not
help.
- Contact
unable to remove the problem.