Cold junction compensation ....................... < ±1.0°C
Sensor error detection ................................ Yes
Sensor error current:
When detecting .................................... Nom. 33 mA
Else....................................................... 0 mA
Voltage input:
Measurement range .................................... -12...150 mV
Min. span ..................................................... 5 mV
Input resistance ........................................... 10 M
Output:
Current output:
Signal range ................................................ 4...20 mA
Min. signal range ......................................... 16 mA
Updating time .............................................. 440 ms
Output signal at EEprom error .................... 3.5 mA
Load resistance ........................................... (V
Load stability .................................................... < ±0.01% af span / 100
Sensor error detection:
Programmable ............................................. 3.5...23 mA
NAMUR NE43 Upscale ............................... 23 mA
NAMUR NE43 Downscale ........................... 3.5 mA
EEx / I.S. approval - 5334B:
KEMA 06ATEX0062 X ..................................
Max. amb. temperature for T1...T4 ............. 85°C
Max. amb. temperature for T5 and T6 ....... 60°C
ATEX, applicable in zone ............................. 0, 1, 2, 20, 21 or 22
Ex / I.S. data:
Signal output / supply, terminal 1 to 2:
U
................................................................. : 30 VDC
i
I
................................................................... : 120 mADC
i
P
.................................................................. : 0.84 W
i
L
.................................................................. : 10 µH
i
C
................................................................. : 1.0 nF
i
Sensor input, terminal 3, 4, 5 and 6:
U
................................................................ : 9.6 VDC
o
I
.................................................................. : 25 mA
o
P
................................................................ : 60 mW
o
L
................................................................. : 33 mH
o
C
................................................................ : 2.4 µF
o
20
All manuals and user guides at all-guides.com
- 7.2) / 0.023
supply
II 1 GD, T80°C...T105°C
EEx ia IIC T6 / T4
Marine approval:
Det Norske Veritas, Ships & Offshore ......... Standard for Certification No. 2.4
GOST R approval:
VNIIM & VNIIFTRI, Cert. no. ........................ See www.prelectroncis.com
Observed authority requirements:
EMC 2004/108/EC ...................................... EN 61326-1
ATEX 94/9/EC .............................................. EN 50014, EN 50020, EN 50284,
Of span = Of the presently selected range
Standard:
IEC 61241-0 and IEC 61241-11
21