Select the type of purge gas, the default is Nitrogen. Argon can be
seleted if needed.
S
Heater Jacket power ON or OFF: setting ON powers on the Heater
Jacket, setting OFF powers it off. The Heater Jacket automatically
switches to OFF when the controller is in the Failure State.
H
S
Heater jacket operating temperature selection. The range is 40-90 °C,
set to 65 °C by default.
T
Detection of the pump's lower bearing temperature. OFF will always
be indicated if there is no sensor on this bearing.
and then:
S
S
Where: XXX = YES or NO.
If YES is selected, the Soft Start mode allows the pump to rump-up
the Normal speed within ten steps.
When NO is selected, the Soft Start mode is deselected and the rump-
up of the pump will be done within 10 minutes.
The controller is factory set to YES.