Télécharger Imprimer la page

Agilent Technologies Turbo-V 2000 ICE Notice De Mode D'emploi page 217

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 37
Connection J7 – Accessories and Options
Interconnections
Figure 12
The 120 Vac, 1 A maximum (independent of line Voltage) output
Voltage is present when the main power switch is set to 1 position
and after START push-button is pressed and will remain present
until a fault condition is displayed on the front panel display or the
turbopump is stopped. The connector J7 is for vent device, forepump
connector coils, pump cooling fan etc. To make connections, remove
the plug and wire the pins (maximum wire size 18 AWG, 1 mm²) as
indicated in the figure to obtain the desired capability.
Forepump relay coil is an independent user supplied item.
NOTE
Turbo-V 2000 ICE Controller User Manual / 87-900-929-01 (L)
J7 connectors
Technical Information
Interconnections
217/260
17