All the logic output from the controller must be connected at P1
mating connector. With the optional P1 mating connector make the
connection with AWG 24 (0.25 mm2) or smaller wire to the pins
indicated in the figure to obtain the desired capability. The following
table describes the signals available on the connector.
The following figure shows a typical logic output connection (relay
coil) but any other device may be connected e.g. a LED, a computer,
etc., and the related simplified circuit of the controller.
Figure 10
Connection J2 - Semi Specifications Connector
Figure 11
Turbo-V 2000 ICE Controller User Manual / 87-900-929-01 (L)
Typical output connection open collector
MALE CONNECTOR
(ANELVA CO NNECTOR)
8
5
7
14
6
J2
13
5
I/O
12
4
11
3
10
2
9
1
Semi Specifications Connector
Technical Information
Interconnections
GND
REMOTE/LOCAL (OPEN CO LLECTOR)
GND
FAILURE (OPEN COLLECTOR)
GND
BRAKING (OPEN COLLECTOR)
GND
NORMAL S1 (OPEN CO LLECTOR)
GND
STARTING S1 (O PEN COLLECTO R)
GND
START S1
17
215/260