17
Technical Information
Use
X
P
Figure 21
Where:
Front / Remote / Serial Selection
Press CYCLE NUMBER and PUMP CURRENT pushbuttons together
for at least 2 seconds and the processor enters a routine where it is
possible to program the controller.
In this routine, the CYCLE push-button is used for choosing/changing
the value or condition; the PUMP CURRENT push-button is used to
enter and confirm the value.
236/260
Press the CYCLE NUMBER push-button twice and the display
shows:
X
X
X
C
Y
U
M
P
CYCLE = are the cycles performed (range 0 to 9999)
m = is the elapsed time related to the cycle number displayed
(range 0 to 99999 minutes)
PUMP LIFE = is the total operation time of the pump (range 0 to
99999 hours).
Turbo-V 2000 ICE Controller User Manual / 87-900-929-01 (L)
C
L
E
L
I
F
E
X
X
X
X
X
X
X
X
X
X
m
h