Controller Controls, Indicators and Connectors
The following paragraph illustrates the Controller control panel and
interconnection panel. More details are contained in the appendix
"Technical Information".
Figure 2
1
Keyboard push-button for HIGH/LOW SPEED mode selection. It is active only when the
front panel operation has been selected. Pressed repeatedly, toggles between HIGH
SPEED and LOW SPEED.
2
Keyboard push-button for START, STOP, RESET mode selection. It is active only when the
front panel operation has been selected. By pressing once the starting phase begins; if
pressed again it stops the pump. If the pump has been stopped automatically by a fault,
this push-button must be pressed once to reset the controller and a second time to
restart the pump.
3
Keyboard push-button to recall on the display the cycle number, cycle time and pumplife.
4
Keyboard push-button to recall on the display the pump current, pump temperature, pump
power, rotational speed, flow value detected, type of gas and alarm status. It is always
active regardless of the operating mode selected. Push-buttons 3 and 4, if pressed
together for at least 2 seconds put the controller in a routine where it is possible to select
some operation parameters.
5
LCD back-lighted alphanumeric display: dot matrix 2 lines x 16 characters.
Turbo-V 2000 ICE Controller User Manual / 87-900-929-01 (L)
Controller 969-9448, 969-9449 front panel
16
Instructions for Use
Use
199/260