Télécharger Imprimer la page

Agilent Technologies Turbo-V 2000 ICE Notice De Mode D'emploi page 231

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 37
WIN
206
207
208
209
210
211
212
213
214
215
300
301
302
400
402
WIN = Window
R = Read
W = Write
T = Type
L = Logical
N = Numeric
A = Alphanumeric
Turbo-V 2000 ICE Controller User Manual / 87-900-929-01 (L)
R
W
T
Description
X
N
ERROR CODE (0÷7) [NO ERROR, OVERVOLTAGE, SHORT
CIRCUIT,CHECK CONNECTION, TOO HIGH LOAD,
OVERRIDE, PUMP OVERTEMP, CONTROLLER OVERTEMP]
X
L
STATE S1 [0=OFF / 1=ON]
X
L
STATE S2 [0=OFF / 1=ON]
X
N
UPPER BEARING TEMPERATURE [0-99 °C]
X
N
LOWER BEARING TEMPERATURE [0-99 °C]
X
N
FLOW READ [0-100 sccm]
X
L
STATUS OUTPUT FLOW METER [0=OFF / 1=ON]
X
N
FLOW METER ALARM TIME RELATING TO THE LAST
EVENT
X
N
FLOW METER ALARM CYCLE NUMBER RELATING TO THE
LAST EVENT
X
N
FLOW METER ALARM EVENT NUMBER
X
N
CYCLE TIME (0-999.999) IN MINUTES
X
N
CYCLE NUMBER (0- 65.535) IN COUNTS
X
N
PUMP LIFE (0-999.999) IN HOURS
X
A
CRC PROGRAM LISTING
X
A
CRC PARAMETER LISTING
Technical Information
Examples
231/260
17

Publicité

loading

Ce manuel est également adapté pour:

969-9448969-9449