For Your Safety
Strictly follow the Instructions for Use
Any use of the apparatus requires a full understanding
and strict observation of these instructions. The
apparatus is only to be used for purposes specified
here.
Maintenance
The apparatus must be inspected and serviced
regularly by suitably qualified persons.
Repairs to the apparatus may only be carried out by
trained service personnel.
We recommend that you enter into a service contract
with Dräger Safety and that all repairs be conducted by
Dräger Safety.
Only authentic Dräger spare parts may be used for
maintenance.
Accessories
Do not use accessory parts other than those listed in
the order list.
Use in areas subject to explosion hazards
Equipment or components which have been inspected
and approved in accordance with national regulations
concerning electrical equipment in explosion-hazard
areas (ElexV) or European explosion-protection
regulations, may be used only under the conditions
explicitly specified in the approval.
The equipment or components may not be modified in
any manner. The use of faulty or incomplete parts is
forbidden.
The appropriate regulations must be observed (e.g.
EIExV and EN 50014) when carrying out repairs on the
equipment or components.
Liability for proper function or damage
The liability for the proper function of the apparatus is
irrevocably transferred to the owner or operator, to the
extent that the apparatus was serviced or repaired by
personnel not employed or authorized by Dräger Safety
or when the apparatus was used in a manner not
conforming to its intended use.
Dräger Safety cannot be held responsible for damage
caused by non-compliance with the recommendations
given above.
The warranty and liability provisions of the terms of sale
and delivery of Dräger Safety are likewise not modified
by the recommendations given above.
Dräger Safety AG & Co. KGaA
8
English
Intended Use
Gas Detector Pump for all Dräger short term tubes and
sampling tubes.
What is What
1
2
3
D
4
5
6
7
accuro
8
1
Tube opener with ceramic cutter
2
Cover plate
3
Socket
4
Rubber bellows
5
Lower pump plate
6
Stroke Counter Reset Button
7
Stroke counter
8
End-of-Stroke Indicator (End-of-Stroke membrane)
9
Bolt to secure lower pump plate in place
10 Hole to break off the tube tip
11 Hole for attachment of loop handle
12 Loop handle with snap hook
Preparation
Pump leak test (A)
Insert unopened Dräger tube into socket (A1).
Squeeze pump completely and release (A2).
Pump is adequately leak-proof if the end-of-stroke
indicator (B) has not appeared after 15 minutes.
If it has, then see chapter "Fault, Cause, Remedy".
Remove unopened Dräger tube from socket.
Press button to reset counter to zero (C);
button is recessed to prevent accidental resetting.
Preparing the tube (D)
9
Select appropriate Dräger tube for the measurement
required. Follow the relevant Instructions for Use!
Open tip with Dräger tube opener
(item no. 64 01 200).
Or:
Insert the tip into the opening of the lower pump plate
(D1); keep it pressed against ceramic edge while
turning it, so that the glass is scored.
Pull the tube over at an angle until the tip breaks off.
Discard broken tip.
Open the other end of tube in the same way.
10
Or:
Turn handle of pump to the side and insert tube tip in
the free hole of the plate (D2).
Tilt tube sideways until the tip breaks off.
Open other end of tube in the same way.
Be careful when opening the Dräger tube as glass
11
splinters may come off! Keep the pump facing
downwards by holding the loop handle, so to prevent
12
glass splinters from entering the pump!
Insert Dräger tube (now open at both ends) into
pump, the arrow must point towards the pump (E).
Caution: the tips of the tube may have sharp edges;
risk of injury!