Making a Measurement
Check number of strokes required from the
Instructions for Use of the Dräger tube.
Hold pump so that the End-of-stroke indicator and the
stroke counter are facing the user (F).
Squeeze pump until stop.
Release pump until bellows are fully expanded.
When the end-of-stroke indicator (F1) appears,
squeeze pump completely again.
Repeat until the number on the stroke counter (F2)
corresponds to that given in the Instructions for Use.
Evaluation of the result in accordance with the
Instructions for Use of the tube in question.
Remove used Dräger tube from socket.
Flush pump with a few pump strokes in clean air.
Maintenance
To clean or replace the exhaust valve
Unscrew bolt by using the hexagonal screwdriver
supplied (G1).
Lift off lower pump plate, e. g. by using a coin. (G2).
Pull valve disc from its seating (H1).
Wipe valve seating with a damp cloth.
Moisten the stem on a new valve disc and push it
firmly into the center hole of the valve seat.
Check correct seating by pulling on valve disc lightly.
Put back lower pump plate and press it into place
(H2).
Insert bolt again and screw down handtight.
Perform the leak test as described (A).
Replacing the socket/
Replacing the membrane
Press spring hook down with a suitable tool
(e. g. ball pen) (I1).
Take out cover plate (I2).
Cover plate holds End-of-Stroke Indicator assembly
in place.
Press spring flanges together and remove socket. (J).
Insert new socket.
Using a new End-of-Stroke Indicator Membrane,
place parts into the housing in the sequence
illustrated (K).
Observe appropriate position.
Insert spring hook of cover plate and push on firmly.
Perform the leak test as described (A).
English
Technical Data
3
Stroke volume
100 ±5 cm
Length
170 mm
Width
45 mm
Height
85 mm
Weight
250 g
Ambient conditions
(Operation and storage)
o
Temperature
–20
C to +50
Humidity
0 to 95 % non condensing
Tests and Approvals
Ex-Approvals
ATEX:
I M1
o
II 1G IIC T6 – 20
C ≤ Ta ≤ + 50
II 1D
BVS 04 ATEX H 068
ATTENTION!
Some Dräger tubes must not be used in
areas subject to explosion hazards.
See the relevant Instructions for Use of the
Dräger tubes.
Fault, Cause, Remedy
Fault
Cause
End-of-stroke
Foreign matter in
indicator visible
exhaust valve,
within 15 min
Defective valve
disc
Tube too loose in
socket
o
C
Pump does not
One of the tube
open
ends has not
been opened
Clogged socket Replace socket,
Order List
o
C
Name
accuro Gas detector pump
Spare part kit
Loop handle, complete
Opener for Dräger tubes
Extension hose 3 m
Gas detector kit
accuro 2000
Remedy
Clean valve and
valve seating.
Replace valve
when required,
(K)
Insert tube
tightly into
socket, (E)
Open tube at
both ends, (D)
(J)
Item no.
64 00 000
64 00 220
64 00 208
64 01 200
64 00 077
64 00 260
64 00 200
9