Sine Arb (Option E36150ADVU)
Step 1– Select Arb waveform
Press Arb to access the Arb selection menu. Then select Sine.
Press Back to return to Arb main menu.
Step 2 – Configure the Sine Arb
Configure the Sine Arb accordingly. Refer to the below table for details.
During configuration, ensure the offset value (V1) must always be greater than the waveform amplitude (V0).
Parameter
Available Key Settings
Arb Type
Voltage or Current
Amplitude (I0 or
minimum to maximum
V0)
Offset (I1 or V1)
minimum to maximum
Frequency (f)
2.7778 μHz - 100 Hz
82
Sets the output Arb type.
Default is Voltage.
Action required:
Press Arb Type to toggle between Volt and Curr.
Volt
Selects voltage as output Arb type.
Curr
Selects current as output Arb type.
Sets the amplitude or peak value.
Sets the offset value.
Sets the frequency of the sine wave.
Description
Keysight E36150 Series User's Guide