Utilities Menu - Test / Setup
Test / Setup provides access to calibration, self-test, as well as configures user preferences, sets the date and time,
installs licenses, and accesses the help:
MEMORY SANITIZATION PROCEDURE
Refer to the E36150 Series Security Guide for instructions to perform a memory sanitization procedure that sanitizes
all user-accessible instrument memory and restarts the instrument.
Calibration
Calibration accesses the instrument calibration procedure.
The default security passcode is 0. Press Login to access the calibration options.
Perform Calibration opens the calibration menu. Make sure you have run the self-test before performing any
calibration. Refer to "Calibration Adjustment Procedures" in the Service Guide for details.
Press Change Passcode to change the passcode using the numeric keypad. The passcode can be set up to 9 digits.
Auto save On will automatically save the calibration data upon logout. When Auto save is Off, press Cal Save to save
the calibration data.
Reset to Factory allows you to reset the power supply to factory state. (Option SEC)