Parameter
Available Key Settings
Trig Src
Key, IO, or Rmt
Repeat Count
1 to 4096
Continuous
-
Step 4 – Run the Arb sequence
When Arb sequence is initiated, all the Arb properties cannot be configured, which includes adding or
removing step from the sequence list.
– Set Trigger Source to Key. Press Trig Src Key.
– Press the color-coded [On] key to enable the output.
– Press Run to start the Arb operation. To abort the operation, press Stop. Alternatively, press [List Run/Stop].
– Press Back to exit and return to the previous menu.
– Press [Scope/Datalog] to enter Scope view and display the Arb.
From the remote interface:
Observe the following when creating or editing an Arb sequence:
– The Arb function type must match the Arb type specified in each sequence step.
– Sequence steps must be specified sequentially. The last value in the parameter list is the sequence step
number.
– When a step is added, all Arb parameters must be entered.
The following commands program a voltage sequence comprised of a pulse, a ramp, and a sine Arb with the pulse
Arb repeated twice.
To set up output 1 to program a sequence of voltage waveforms:
ARB:FUNC:TYPE VOLT
ARB:FUNC:SHAP SEQ
ARB:SEQ:RESet
Keysight E36150 Series User's Guide
Sets the trigger source for the system.
Default is Rmt.
Action required:
Press Trig Src to toggle between Key, IO, and Rmt.
Key (Arb Run/Stop Key)
IO (DIO Trigger In)
Rmt (Remote Command)
Sets the Arb sequence repeat count. This sets the number of times that the Arb sequence is
executed before it completes.
Default is 1.
Enables check box to repeat the sequence step continuously.
Description
Selects the Arb Run Stop key or [List Run/Stop] as
a trigger source.
Selects any configured digital IO (with Trigger
Input function) as a trigger source
Selects a remote interface command as a trigger
source.
105