Alimentation cc à sélection automatique

Publicité

Les langues disponibles

Les langues disponibles

Scope View
Press [Scope/Datalog] key to view the scope. This key toggles between the Scope view and Data Logger view.
Items
1
Trace controls
2
Data bar
3
Time/Div
4
Offset time
5
Arb status
6
Trigger mode
7
Trigger Source
Amplitude
8
Sample period
9
Range
10
Scope status
Keysight E36150 Series User's Guide
Identifies the voltage, current , or power that will be displayed. √ indicates the trace is on. Dashes (---
-) indicate the trace is off. Select the trace and press [Enter] to turn it on or off.
The data bar represents all of the logged data. The yellow part indicates the portion of the data that is
visible on the display. The black part represents the data that is not visible.
Identifies the horizontal time-base setting. This can be adjusted using the Horizontal knob after press-
ing Knob Sc.
Indicates the time from the trigger point indicator to the vertical center line of the grid. Negative
values indicate the center line is to the left of the trigger point. Positive values indicate the center line
is to the right of the trigger point. The trigger point can be adjusted using the Horizontal knob after
pressing Knob Ofs.
Indicates the Arb is running. No indicator when Arb is idle.
Indicates the trigger mode (Auto, Single, or Triggered).
In the figure, the trigger source is a voltage level on output 1.
Up trigger indicates the measurement will be triggered on the up-slope (positive).
Down trigger indicates the measurement will be triggered on the down-slope (negative).
If the trigger source is set to a voltage or current level, the amplitude of the trigger level is indicated
below the trigger source. In the figure, the voltage trigger level is set to 0 V.
The indicated scope sample period is based on the horizontal time/div. setting. When the time/div.
setting is less than 2 ms/division, the scope will sample at its fastest rate, 10 µs.
Range indicates the measurement range setting of the selected trace.
Indicates whether the scope is Running, Stopped, or Waiting for a trigger.
Description
133