Power-Off is activated, if the symbol
if the rotary switch 5 is switched on again from switch position „OFF". Automatic switch-
off can be deactivated by pressing the HOLD key and by simultaneously switching on
the BENNING CM 9-1 from the switching position "OFF". The symbol
from the display 9 .
5.1.12 The BENNING CM 9-1 is supplied by two 1.5 V batteries (IEC LR03/ AAA/ micro).
5.1.13 If the battery voltage drops below the specified operating voltage of the BENNING
CM 9-1, then a battery symbol
5.1.14 The battery life depends on the measuring function used and is approximately 40 hours
to 60 hours without using the acoustic continuity test and background lighting (alkaline
battery).
5.1.15 Temperature coefficient of the measured value: 0.1 x (stated measuring accuracy)/ °C
< 18 °C or > 28 °C, related to the value for the reference temperature of 23 °C
5.1.16 Dimensions of unit (length x width x height) = 210 x 76 x 33,5 mm.
Weight of unit: 296 g (incl. batteries)
5.1.17 The safety measuring leads and measurement tips supplied are especially suited to the
rated voltage of the BENNING CM 9-1. The measuring tips can be protected by caps.
5.1.18 Widest prong opening: 23 mm
6.

Ambient conditions

-
The BENNING CM 9-1 is intended for making measurements in dry environment.
-
Maximum barometric elevation for making measurements: 2000 m,
- Overvoltage category: IEC 60664/ IEC 61010 → 300 V category IV; 600 V category III
-
Operating class of current sensor: EN 61557-13, class 2, ≤ 30 A/m, @ In: 3.5 mA to 600 mA,
fn: 40 Hz to 1 kHz
-
Contamination class: 2 (EN 61010-1),
-
Protection class: IP 30 (DIN VDE 0470-1, IEC/ EN 60529)
IP 30 means: Protection against access to dangerous parts and protection against solid
impurities of a diameter > 2.5 mm, (3 - first index). No protection against water, (0 - second
index).
-
Operating temperature and relative humidity:
For operating temperatures from 0 °C to 40 °C: relative air humidity lower than 80 %, non-
condensing.
-
Storage temperature: The BENNING CM 9-1 can be stored at temperatures between - 10 °C
and + 60 °C, at a relative air humidity lower than 80 % without batteries.
7.

Electrical specifications

Note: The measuring precision is specified as the sum of
-
a relative fraction of the measured value and
-
a number of digits (counting steps of the least significant digit).
This specified measuring precision is valid for temperatures in the range from 23 °C ± 5 °C and
relative humidity less than 80 %. The stated accuracy is specified for 1 % to 100 % of the final
measuring range value.
7.1 DC voltage ranges
The input resistance is ≥ 2 MΩ.
Measuring range
60.00 V
60.0 V - 600.0 V
7.2 AC voltage range
The measured value is obtained and displayed as real r.m.s. value (True RMS, AC coupling).
Its calibration is adapted to sinusoidal curves. In case of deviations from this curve shape, the
accuracy of the displayed value decreases.
Crest factor < 2.0 up to 100 % of each final measuring range value
Crest factor < 4.0 up to 50 % of each final measuring range value
The input resistance is ≥ 2 MΩ.
Measuring range
60.00 V
60.0 V - 600.0 V
10/ 2020
appears in the display 9 .
Resolution
0.01 V
0.1 V
Resolution
Measurement accuracy (15 Hz - 1 kHz)
0.01 V
0.1 V
BENNING CM 9-1
is shown on the display 9 ). It switches on again,
Measurement accuracy
± (1.0 % + 4 digits)
± (1.0 % + 4 digits)
± (1.2 % + 5 digits)
± (1.2 % + 5 digits)
disappears
Overload protection
600 V AC/DC
600 V AC/DC
Overload protection
600 V AC/DC
600 V AC/DC