7.3 AC current range
The measured value is obtained and displayed as real r.m.s. value (True RMS, AC coupling).
Its calibration is adapted to sinusoidal curves. In case of deviations from this curve shape, the
accuracy of the displayed value decreases.
Crest factor < 2.0 up to 100 % of each final measuring range value
Crest factor < 4.0 up to 50 % of each final measuring range value
Overload protection: 60 A AC/DC
low-pass filter (40 Hz - 70 Hz) deactivated
Measuring range
6.000 mA
*1
60.00 mA
600.0 mA
6.000 A
60.00 A
Lowest value displayed: 0.010 mA
*1
Frequency characteristics according to EN 61557-16 (VDE 0413-16)
Limiting frequency fg (- 3 dB): approx. 1 kHz
low-pass filter (40 Hz - 70 Hz) activated
Measuring range
6.000 mA
*1
60.00 mA
600.0 mA
6.000 A
60.00 A
Lowest value displayed: 0.010 mA
*1
Limiting frequency fg (- 3 dB): approx. 180 Hz
7.4 Resistance range
Open circuit-voltage: approx. 3 V, max. testing current 1 mA
Measuring range
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
7.5 Continuity test
Open circuit-voltage: approx. 3 V, max. testing current 1 mA
The built-in buzzer sounds when resistance is less than 45 Ω.
7.6 Variations and uncertainties
Variation
E1 Position
E2 Supply voltage
E3 Temperature
E9 Distorted curve shape
E10 DC components in the mains
10/ 2020
Resolution
15 Hz ~ 40 Hz
0.001 mA
± (2.0 % + 10 digits)
0.01 mA
0.1 mA
± (2.0 % + 7 digits)
0.001 A
0.01 A
Resolution
0.001 mA
0.01 mA
0.1 mA
0.001 A
0.01 A
Resolution
0.1 Ω
1 Ω
10 Ω
100 Ω
BENNING CM 9-1
Measurement accuracy
± (1.0 % + 10 digits)
± (1.0 % + 7 digits)
Measurement accuracy
± (1.0 % + 10 digits)
± (1.0 % + 7 digits)
Measurement accuracy
± (1.0 % + 4 digits)
1 % of the measured value
-
0.1 x (stated measuring accuracy)/ °C
(<18 °C or >28 °C)
-
-
40 Hz ~ 70 Hz
± (2.5 % + 10 digits)
± (2.5 % + 7 digits)
40 Hz ~ 70 Hz
Overload protection
600 V AC/DC
70 Hz ~ 200 Hz
23