After linearisation the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will return
into weighing mode automatically.
6.9.2
Non-verified weighing systems
Call-up menu item P3 CALCalLiner, see chap. 6.9.1
Confirm by
, the password query „Pn" will be displayed.
Press
,
,
Ensure that there are no objects on the weighing pan.
Wait for stability display, then press
When "Ld 1" is displayed, put the first adjustment weight (1/3
max) carefully in the centre of the weighing platform. Wait for
stability display, then press
When "Ld 2" is displayed, put the second adjustment weight
(2/3 max) carefully in the centre of the weighing platform.
Wait for stability display, then press
When "Ld 3" is displayed, put the third adjustment weight
(max) carefully in the centre of the weighing platform. Wait
for stability display, then press
After a successful linearisation the balance will carry out a
self-test. Remove adjusting weight during selftest, balance
will return into weighing mode automatically.
20
or
,
,
subsequently.
.
.
.
.
KFB/KFN-TAM-BA_IA-e-1832