Menu (menu)
5.5
Adjustment mode (Calib. Mode)
Off
Adjust measuring cell with accompanying density standard or distilled
water.
On
Adjust measuring cell with desired density standard.
The following input is required:
Density
Temp
α
1000
x
5.6
Interface
PRN
Printer interface
Data transfer to the printer
Printer with serial interface and connected infrared adapter. Results are
formatted for output to a printer.
Transfer rate (baud rate), parity and stop and data bits must be config-
ured according to the peripheral device.
The following settings are required for the METTLER TOLEDO LC-P45
printer:
Baud rate
9600
Parity
none
Stop bits
1
Data bits
8
RS
Serial interface, data transfer to the PC via the infrared adapter.
Transfer rate (baud rate), parity and stop and data bits must be config-
ured as in the MiniTerm program, see Chapter 8.
IrDA
Data transfer to the PC with the integrated IrDA interface in accordance
with protocol 1.20.
5.7
Beep
Off
Beep off.
On
Beep on.
5.8
LCD-Contrast
Display contrast adjustable to one of 9 levels with the
5.9
Automatic switch-off (Power)
Off
Automatic switch-off off. The instrument must be switched off manually.
On
The instrument switches off automatically if not operated for 10 min-
utes.
5.10
Software version (Version No.)
The software version is displayed.
12
Density of the density standard (g/cm
reference temperature.
Reference temperature
Temperature-compensation coefficient of the
density standard
1000
•
METTLER TOLEDO Densito 30P
3
) at the
keys.