Shower and Toilet Commode Ocean E-VIP
Fault
The seat frame will not move
upwards
or
The seat frame will not move
downwards
or
The seat frame will not tilt
Battery will not charge
Implementing a reset
1. Hold down the following three buttons simultaneously for about
three seconds on the hand control: Red STOP button E, Forward
button C and Up button H.
An acoustic signal will be heard while the three buttons are held down.
2. When „SERVICE" indicator lamp F goes out, release the
buttons. The device is ready to use again.
3. Then move the drives for adjusting the height to the lowest positi-
on until you hear the drives switch off.
4. Move the tilt drive to the rear end position until you hear the drive
switch off. (The red LED flashes and an acoustic signal is heard).
!
IMPORTANT
▸ If the fault recurs after the reset, replace the device or the
faulty parts and send them to us.
20
Possible cause(s)
Weight limit exceeded
Cable not correctly inserted
Connector strip not connected
Cable incorrectly attached to controller
Connector cable broken
Connector cable not fully inserted
Controller faulty
Battery flat
Hand control cable is not inserted correctly
Hand control faulty
Faulty battery
Connector not correctly inserted
Remedy
▸ Max. load 150 kg
▸ Insert the cable correctly
▸ Connector strip connected (engaged)
▸ Motor for tilting always on channel 7
▸ Replace the cable
▸ Insert the cable fully and close the safety bar.
▸ Implement reset (see below)
▸ Charge the battery (see Section 4.11) or use a replace-
ment battery
▸ Insert the cable fully
▸ Implement reset (see below)
or
▸ Replace the hand control
▸ Replace the battery
▸ Check that all the connectors are positioned correctly
10 Technical data
E
H
K
F
L
B
Technical data
I
C
J
G
D
A