8.2
Servicing Dentsply Sirona quick coupling
8.2.1
Replacing O-rings
1. Remove the defective O-ring.
Using T1 Spray
A
2. Slide the tool (A), with the new O-ring attached, to just in front of the
3. Insert the O-ring. Do not use any sharp objects.
4. Lightly oil the O-ring with Midwest
B
corresponding groove (B).
NOTICE! Do not use Vaseline or silicone grease on the O-rings.
8 Maintenance
®
Plus Aerosol Spray.
59