7 Conditioning
54
All manuals and user guides at all-guides.com
7.5
Sterilizing
Intervals:
● Prior to initial operation
● Prior to every other use
Process
✔ The high-speed handpiece is cleaned and disinfected.
✔ High-speed handpiece maintenance is complete.
✔ If necessary, the high-speed handpiece can be sterilized in
packaging suitable for sterilization and storage, e.g. paper/plastic
composite packaging or container.
➢ Sterilize the high-speed handpiece in the steam sterilizer with
saturated water vapor.
Operating Instructions Midwest Stylus
®
®
Plus / Midwest
Phoenix ™ /Pro
6607290 D3673.201.01.03.02