Operating Instructions Midwest Stylus
6607290 D3673.201.01.03.02
®
®
Plus / Midwest
Phoenix ™ /Pro
All manuals and user guides at all-guides.com
6
Follow-up
6.1
After each treatment session
NOTICE! Condition immediately, or at the latest, one hour after
treatment.
NOTICE! Only use Midwest
✔ The bur must not be moving.
✔ Wear appropriate protective clothing.
1. Purge the water and air channels on the treatment center for 30
seconds.
2. Remove the bur with tweezers.
3. Predisinfect directly at the treatment center [ → 43].
4. Remove the high-speed handpiece.
5. Transport the high-speed handpiece to the hygiene room in a
suitable transport container.
®
Plus Aerosol Spray.
6 Follow-up
41