Lower anterior teeth are
severely protruded
SUS³ telescope rod
catches / rubs during mouth
closure
Fixing screw on the SUS³ arch
adapter comes loose
* The Dentaurum version of this prescription is not claimed to be a duplication of any other, nor does Dentaurum imply that it is
endorsed in any way by Drs. McLaughlin, Bennett and Trevisi.
Too little anchorage and/or
overactivation
SUS³ telescope rod is bent after
uncontrolled mouth closure (e.g.
after yawning)
Fixing screw too loose or not
tightened enough
- Insert a strong rectangular stainless steel arch, min. 0.43 x 0.56 mm /
17 x 22 in slot 18 or 0.53 X 0.64 mm / 21 X 25 in slot 22 with
lingual crown torque, McLaughlin-Bennett-Trevisi* brackets or
low torque brackets and bend it distally
- SUPER Elasto-Force plastic chain (power chain) from 6 – 6
- Temporary anchorage with tomas
Replace the SUS³ telescope rod (can be ordered separately)
Tighten the fixing screw and fix in position using a screw securing
adhesive (e.g. Ceka
Bond)
®
EN
-pin
®
19