en
12
3
Product description
The label on the housing contains the following information
METTLER TOLEDO
www.mt.com
Im Hackacker 15
CH-8902 Urdorf, Switzerland
P/N
S/N
PS
TS
CE
xxxx
See instruction manual
4
Installation and commissioning
1. While unpacking, check the housing for mechanical
damage. Please immediately report any damage to your
local METTLER TOLEDO representative.
2. Ensure the housing is fully retracted and in the "service"
position.
3. Isolate the pipe / tank / vessel etc. from the housing
(depressurize, empty, rinse, purge and vent, etc.) before
starting any installation work.
4. Fit the housing to flange / NPT thread of process
connection.
5. Attach the flushing lines (optional flushing chamber
when applicable).
6. Fit the electrode / sensor to the housing and connect the
electrode / sensor and the transmitter with corresponding
cable.
7. Loosen fixing screw <3> to release the immersion tube
<1> by hand or using a no. 38 spanner.
8. Open the ball valve <2>.
9. Slide the immersion tube into the measuring position as
far as necessary or until it stops.
10. If the housing is not inserted as far as the stop, fix the
adjusting ring <4> in the desired position.
11. Tighten fixing screw <3> by using a no. 38 spanner.
Refer to housing instruction manual for detailed description.
InTrac 785 / 787
30 313 391 C
1
Manufacturer of the
housing and contact
d etails
Part number
Serial number
Maximum permissible
pressure
Maximum permissible
temperature
CE marking with notified
body number
See instruction manual
for detailed description
4
3
2
© 11 / 2022 METTLER TOLEDO
Printed in Switzerland
InTrac 785 / 787