4.
Select X-ray intensity.
<Console Software>
5.
The values of tube voltage and current are configured automatically according to the
patient's gender / age group and X-ray intensity. Click the UP/DOWN arrow to adjust
kVp and mA. The dose is adjustable by ±1 kVp and ± 0.1 mA respectively.
<Console Software>
6.
Click CONFIRM button when exposure parameter setting is completed.
<Console Software>
7.
Bring the Plaster Cast to the equipment.
Green16 User Manual
<Touch Screen>
When you click CONFIRM button,
▪ The Rotating Unit will move to its initial scanning position.
▪ The Vertical Beam will be activated to make patient positioning
easier.
▪ The DAP (Dose Area Product), Scan Time and Exposure Time will
be displayed below Patient Information window.
6. Acquiring 3D MODEL Scan Images
<Touch Screen>
<Touch Screen>
83