3. Acquiring PANO Images
Laser Beam Aligning
Align the Vertical Beam with the center of the face (Mid-sagittal Line). (It's to prevent
1.
horizontal expansion of the image)
2.
Align the Horizontal Beam in a straight line to the Frankfurt Line on the patient's face.
Use the Horizontal Beam button on the control panel to position it. Make sure that the
Horizontal Beam is aligned to the patient's face horizontally.
40
Be careful not to shine the laser beam directly into the person's eyes.
Doing so may result in vision loss.
If the laser beam is not correctly positioned there may be distortion,
causing the image to be enlarged or reduced, or ghost shadows may
occur and lower the image quality. Be sure to align the laser beam
properly.