The values of tube voltage and current are configured automatically according to the
patient's gender / age group and X-ray intensity. Click the UP/DOWN arrow to adjust
kVp and mA. The dose is adjustable by ±1 kVp and ± 0.1 mA respectively.
<Console Software>
8.
Click CONFIRM button when exposure parameter setting is completed.
<Console Software>
9.
Guide the patient to the equipment.
Green16 User Manual
<Touch Screen>
When you click CONFIRM button,
▪ The Rotating Unit will move to its initial scanning position.
▪ The Vertical Beam will be activated to make patient positioning
easier.
▪ The DAP (Dose Area Product), Scan Time and Exposure Time will
be displayed below Patient Information window.
5. Acquiring CBCT Images
<Touch Screen>
71