Checking that the contents of the calibration counter are correct
This seca scale is calibrated. Calibration may be performed only by authorised centres.
To ensure this, the scale is fitted with a calibration counter which records any change to
data relevant to calibration. If you want to check whether the scale is properly calibrated,
proceed as follows:
– If necessary, switch off the scale.
– Keep any key depressed and start the
scale.
– The current contents of the calibration
counter are shown on the display for a
few seconds.
– Compare the output contents of the cal-
ibration counter with the number shown
on the calibration counter sticker.
For valid calibration, both numbers must
match. If sticker and calibration counter do
not match, then re-calibration is required.
Contact your service agent or seca Cus-
tomer Service.
If a supplementary calibration is necessary,
then the supplementary calibration mark il-
lustrated opposite is used to record the
calibration counter position instead of the
seca calibration mark shown above. This
mark is secured with an additional seal by
the person authorised to perform the sup-
plementary calibration. The supplementary
calibration mark is available from seca
Customer Service under the number
14-05-01-886.
6. Cleaning
Clean the tray and the housing as required with a domestic cleaning agent or commercially-
available disinfectant. Follow the manufacturer's instructions.
Model 375/376
Both numbers must match.